12243413 |
Wang R, Liu Z, Cai R, Li X: A new spectrofluorometric method for the determination of ascorbic acid based on its activating effect on a hemoglobin-catalyzed reaction. Anal Sci. 2002 Sep;18(9):977-80. A spectrofluorometric method for the determination of ascorbic acid (AA) based on its activation on the hemoglobin-catalyzed reaction was proposed. The fluorescence intensity of the product was measured under the optimal experimental conditions, i.e. 4.0 x 10 (-6) M H2O2, 6.0 x 10 (-5) M p-cresol, 1.2 M NH3-NH4Cl (pH 10.4) and 2.0 x 10 (-7) M hemoglobin. The order of additions of the reagents was also studied. The activation of AA was found to be associated with a high ammonia concentration. The linear range of the method was 9.0 x 10 (-10)-3.6 x 10 (-8) M of AA. The detection limit was calculated to be 3.0 x 10 (-10) M. The relative standard deviation of this method is 1.6% at 7.0 x 10 (-9) M for 11 determinations. |
3(0,0,0,3) |