11148184 |
Buscarlet L, Volland H, Dupret-Carruel J, Jolivet M, Grassi J, Creminon C, Taran F, Pradelles P: Use of free radical chemistry in an immunometric assay for 17 beta-estradiol. Clin Chem. 2001 Jan;47(1):102-9. BACKGROUND: We wished to develop an enzyme immunometric assay for 17 beta-estradiol (E2) in human serum using solid-phase immobilized epitope immunoassay (SPIE-IA) technology and free radical chemistry. METHODS: We used an anti-estradiol monoclonal antibody as capture antibody and Fenton-like reagents to cross-link it to E2. The same antibody, labeled with acetylcholinesterase, was used for detection. Serum was diluted 10-fold before assay. RESULTS: After correction by the dilution factor, the detection limit was 5 ng/L for human serum and intra- and interassay CVs were <7% and 15%, respectively, at concentrations of 169-2845 ng/L. No cross-reactivity was seen with other natural steroids. In comparison with a competitive commercial RIA performed on 88 undiluted human sera, the slope (SD) of the regression line was 1.05 (+/- 0.02) and the intercept was 47 (+/-27) ng/L (S (y/x) = 186 ng/L) at concentrations of 20-5000 ng/L (r (2) = 0.97). CONCLUSIONS: The use of Fenton-like chemistry in SPIE-IA technology allows a sensitive measurement of E2 in human serum and could be a new approach for the development of sensitive immunoassays. |
1(0,0,0,1) |